dc.contributor.author |
Bansal, Aayushi |
|
dc.date.accessioned |
2023-05-29T06:28:57Z |
|
dc.date.available |
2023-05-29T06:28:57Z |
|
dc.date.issued |
2022 |
|
dc.identifier.citation |
Bansal Auyushi, 2022, Determination of ETORICOXIB in Bulk Formulations by RP-HPLC methods, World Journal of Pharmaceutical Research, Volume 11, Issue 15, 1128-1136. , ISSN 2277– 7105, |
en_US |
dc.identifier.issn |
2277– 7105 |
|
dc.identifier.uri |
http://10.9.150.37:8080/dspace//handle/atmiyauni/1160 |
|
dc.description |
I am deeply indebted to Akums Drugs and Pharmaceuticals Ltd. for letting me to conduct my
research work. Words cannot express my appreciation to my family and friends for their
constant support and help. |
en_US |
dc.description.abstract |
For the determination of Etoricoxib in bulk formulations a rapid,
selective and simple RP-HPLC method was formed and used, by
optimizing stationary phase, composition of mobile phase and other
parameters. Study was performed on RP- C-18 (250 x 4.6 mm), 5 µm
column (Symmetry, Waters) with the mobile phase Acetonitrile:
Phosphate buffer in the ratio of 70:30 v/v. At a flow rate of 1 ml/ min,
the mobile phase was pumped and the detection was monitored at 220
nm. Short retention time was observed of the drug i.e. 9.589 min. The
calibration curve was linear from 70ppm to 130 ppm and correlation
coefficient of 0.998144. The intra-day precision (%RSD) was 0.99 and inter-day precision (%
RSD) was in the range 0.99 to 1.21. The assay was found to be 99.28 %. The above method
was successfully used for the assay of etoricoxib in bulk formulations |
en_US |
dc.language.iso |
en |
en_US |
dc.publisher |
World Journal of Pharmaceutical Research |
en_US |
dc.subject |
Etoricoxib |
en_US |
dc.subject |
RP-HPLC |
en_US |
dc.subject |
Assay |
en_US |
dc.subject |
Bulk formulation |
en_US |
dc.title |
Determination of ETORICOXIB in Bulk Formulations by RP-HPLC methods |
en_US |
dc.type |
Article |
en_US |